Optimizing the Architecture of a Dynamic Spinal Implant for Customized Mechanical Behavior
Langue
EN
Article de revue
Ce document a été publié dans
Procedia CIRP. 2017-06-21, vol. 65, p. 273-278
Résumé en anglais
Non-fusion technology in spine surgery reduces surgical morbidity and degeneration of the adjacent levels by the insertion of dynamic spinal implants. Despite these advantages, a dynamic spinal implant (DSI) generates ...Lire la suite >
Non-fusion technology in spine surgery reduces surgical morbidity and degeneration of the adjacent levels by the insertion of dynamic spinal implants. Despite these advantages, a dynamic spinal implant (DSI) generates complications which require clinical follow-up, the continuous development of constructive solutions and structured optimization of the implant architecture using current mechanical design methods.
This study structures this optimization process of a DSI concept by incorporating the mechanical behavior of the device, design variables and functional requirements into a global design model. The geometric (descriptive anatomy) and mechanical (materials, components, etc.) characteristics are obtained from a literature review. By combining these parameters, variables and requirements, appropriate values can be determined. The resulting mathematical model is then used to design and implement a device that is suitably adapted in movements and stiffness. The model assumes linear or non-linear behavior. We describe the optimization of the design variables to ensure the correct functioning of the mechanism when adapted to the patient. The optimization purpose is to determine the architecture of the implant, the choice of materials and the geometric parameters of implantation. An optimized implant model corresponding to specific degrees of degeneration in the intervertebral joint can then be envisaged.< Réduire
Mots clés en anglais
Dynamic implant
Design
Mechanical Model
Optimization
Unités de recherche