How to assess depth-of-field extension strategies applied to single-molecule super-resolution microscopes?
Langue
en
Communication dans un congrès avec actes
Ce document a été publié dans
Computational Optical Sensing and Imaging, 2021-07-19, Washington. p. CF4B.5
OSA
Résumé en anglais
Phase masks used to extend the depth-of-field (DoF) of localization microscopes are often designed to increase localization accuracy despite a loss in detection probability. We propose a method to optimize DoF extended ...Lire la suite >
Phase masks used to extend the depth-of-field (DoF) of localization microscopes are often designed to increase localization accuracy despite a loss in detection probability. We propose a method to optimize DoF extended single-molecule localization techniques taking into account the trade-off between detection probability and localization accuracy.< Réduire
Origine
Importé de halUnités de recherche