Magnetic field enhancement for femtosecond-laser-ablation mass spectrometry in ambient environments
Langue
en
Article de revue
Ce document a été publié dans
Journal of Analytical Atomic Spectrometry. 2015, vol. 30, n° 11, p. 2303-2306
Royal Society of Chemistry
Résumé en anglais
To improve the sensitivity of open-air femtosecond-laser-ablation mass spectrometry (fs-LA-MS), an external magnetic field was introduced to confine the propagation of the plasma, reduce the plasma neutralization rates, ...Lire la suite >
To improve the sensitivity of open-air femtosecond-laser-ablation mass spectrometry (fs-LA-MS), an external magnetic field was introduced to confine the propagation of the plasma, reduce the plasma neutralization rates, and drive charged particles towards the MS orifice. Signal-to-noise ratios (SNRs) of trace elements, including Al, Si and Pb, in standard NIST samples were obviously improved. The SNR enhancement depends strongly on the direction of the magnetic fields. The maximum enhancement factors were obtained when directing the magnetic fields towards the orifice of the mass spectrometer. The enhancement factors (EFs) of 8.30, 8.22 and 8.50 were achieved for 206Pb, 207Pb and 208Pb in pure lead. The limits of detection (LODs) of 206Pb, 207Pb and 208Pb were lowered from 9.75, 32.9 and 8.19 ppm to 2.32, 6.02 and 1.98 ppm, respectively.< Réduire
Origine
Importé de halUnités de recherche