dc.description.abstractEnBACKGROUND: The accuracy of currently available devices using pulse contour analysis without external calibration for cardiac index (CI) estimation is negatively impacted by hyperdynamic states, low systemic vascular resistance (SVR), and abrupt changes in SVR. The aim of this study was to evaluate the accuracy of a new device, the Pulsioflex (Pulsion Medical System), in patients undergoing liver transplantation. METHODS: Thirty consecutive patients scheduled for liver transplantation were included. CI was monitored using pulmonary arterial catheter (CI-PAC) and Pulsioflex (CI-Pulsio). Simultaneous CI measurements were made intraoperatively at 9 different stages of the procedure. RESULTS: Two hundred seventy pairs of measurements were analyzed. The median CI-Pulsio values (3.3; interquartile range, 2.8-3.8 L·min·m) were significantly different from the median CI-PAC (4.1; interquartile range, 3.1-5.0 L·min·m; P \textless .0001). Bland and Altman analysis showed a mean bias of 0.8 L·min·m and 95% limit of agreement from -2.5 to 4.1 L·min·m. Percentage error was 65% (95% confidence interval, 60%-71%). Considering the variations in CI between 2 stages, the comparison between changes in CI-PAC and changes in CI-Pulsio showed a mean bias of 0.1 L·min·m and 95% limit of agreement of -2.1 to 2.2 L·min·m. When excluding changes in CI \textless0.5 L·min·m (154 paired analyzed), the concordance rate was 62% (95% confidence interval, 54%-70%). The bias between CI-PAC and CI-Pulsio was negatively correlated with SVR (r = -0.67, P \textless .0001). The bias between changes in CI-PAC and changes in CI-Pulsio was also negatively correlated with changes in SVR (r = -0.52, P \textless .0001). CONCLUSIONS: In patients undergoing liver transplantation, Pulsioflex does not accurately estimate CI. Its accuracy is highly impacted by SVR, and it is not able to track changes in CI when large variations in SVR occur.
dc.language.isoENen_US
dc.subject*Article CLINIQUE
dc.title.enImpact of Systemic Vascular Resistance on the Accuracy of the Pulsioflex Device
dc.typeArticle de revueen_US
dc.identifier.doi10.1213/ANE.0000000000001591en_US
dc.subject.halSciences du Vivant [q-bio]/Médecine humaine et pathologieen_US
bordeaux.journalAnesthesia Analgesiaen_US
bordeaux.page487–493en_US
bordeaux.volume124en_US
bordeaux.hal.laboratoriesBiologie des maladies cardiovasculaires - U1034en_US
bordeaux.issue2en_US
bordeaux.institutionUniversité de Bordeauxen_US
bordeaux.peerReviewedouien_US
bordeaux.inpressnonen_US
hal.exportfalse
bordeaux.COinSctx_ver=Z39.88-2004&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.jtitle=Anesthesia%20Analgesia&rft.date=2017&rft.volume=124&rft.issue=2&rft.spage=487%E2%80%93493&rft.epage=487%E2%80%93493&rft.au=BIAIS,%20Matthieu&MAZOCKY,%20Elodie&STECKEN,%20Laurent&PEREIRA,%20Bruno&SESAY,%20Musa&rft.genre=article


Fichier(s) constituant ce document

FichiersTailleFormatVue

Il n'y a pas de fichiers associés à ce document.

Ce document figure dans la(les) collection(s) suivante(s)

Afficher la notice abrégée