Calibration of imaging plate detectors to mono-energetic protons in the range 1-200 MeV
hal.structure.identifier | Centre d'Etudes Lasers Intenses et Applications [CELIA] | |
dc.contributor.author | RABHI, N. | |
hal.structure.identifier | Centre d'Etudes Lasers Intenses et Applications [CELIA] | |
dc.contributor.author | BATANI, D. | |
hal.structure.identifier | Centre d'Etudes Lasers Intenses et Applications [CELIA] | |
dc.contributor.author | BOUTOUX, G. | |
hal.structure.identifier | Centre d'Etudes Lasers Intenses et Applications [CELIA] | |
dc.contributor.author | DUCRET, J.-E. | |
hal.structure.identifier | Centre d'Etudes Lasers Intenses et Applications [CELIA] | |
dc.contributor.author | JAKUBOWSKA, K. | |
hal.structure.identifier | Direction des Applications Militaires [DAM] | |
dc.contributor.author | LANTUEJOUL-THFOIN, I. | |
hal.structure.identifier | Institut Curie [Paris] | |
dc.contributor.author | NAURAYE, C. | |
hal.structure.identifier | Institut Curie [Paris] | |
dc.contributor.author | PATRIARCA, A. | |
hal.structure.identifier | Institut de Physique Nucléaire d'Orsay [IPNO] | |
dc.contributor.author | SAÏD, A. | |
hal.structure.identifier | Institut de Physique Nucléaire d'Orsay [IPNO] | |
dc.contributor.author | SEMSOUM, A. | |
hal.structure.identifier | Centre d'Etudes Nucléaires de Bordeaux Gradignan [CENBG] | |
dc.contributor.author | SERANI, L. | |
hal.structure.identifier | Centre d'Etudes Nucléaires de Bordeaux Gradignan [CENBG] | |
dc.contributor.author | THOMAS, B. | |
hal.structure.identifier | Direction des Applications Militaires [DAM] | |
dc.contributor.author | VAUZOUR, B. | |
dc.date.issued | 2017 | |
dc.description.abstractEn | Responses of Fuji Imaging Plates (IPs) to proton have been measured in the range 1-200 MeV. Mono-energetic protons were produced with the 15 MV ALTO-Tandem accelerator of the Institute of Nuclear Physics (Orsay, France) and, at higher energies, with the 200-MeV isochronous cyclotron of the Institut Curie—Centre de Protonthérapie d’Orsay (Orsay, France). The experimental setups are described and the measured photo-stimulated luminescence responses for MS, SR, and TR IPs are presented and compared to existing data. For the interpretation of the results, a sensitivity model based on the Monte Carlo GEANT4 code has been developed. It enables the calculation of the response functions in a large energy range, from 0.1 to 200 MeV. Finally, we show that our model reproduces accurately the response of more complex detectors, i.e., stack of high-Z filters and IPs, which could be of great interest for diagnostics of Petawatt laser accelerated particles | |
dc.language.iso | en | |
dc.title.en | Calibration of imaging plate detectors to mono-energetic protons in the range 1-200 MeV | |
dc.type | Article de revue | |
dc.identifier.doi | 10.1063/1.5009472 | |
dc.subject.hal | Physique [physics]/Physique [physics]/Instrumentations et Détecteurs [physics.ins-det] | |
bordeaux.journal | Rev.Sci.Instrum. | |
bordeaux.page | 113301 | |
bordeaux.volume | 88 | |
bordeaux.issue | 11 | |
bordeaux.peerReviewed | oui | |
hal.identifier | hal-01851242 | |
hal.version | 1 | |
hal.popular | non | |
hal.audience | Internationale | |
hal.origin.link | https://hal.archives-ouvertes.fr//hal-01851242v1 | |
bordeaux.COinS | ctx_ver=Z39.88-2004&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.jtitle=Rev.Sci.Instrum.&rft.date=2017&rft.volume=88&rft.issue=11&rft.spage=113301&rft.epage=113301&rft.au=RABHI,%20N.&BATANI,%20D.&BOUTOUX,%20G.&DUCRET,%20J.-E.&JAKUBOWSKA,%20K.&rft.genre=article |
Fichier(s) constituant ce document
Fichiers | Taille | Format | Vue |
---|---|---|---|
Il n'y a pas de fichiers associés à ce document. |