Lumbar spine posterior corner detection in X-rays using Haar-based features
Idioma
en
Communication dans un congrès avec actes
Este ítem está publicado en
2016 IEEE 13th International Symposium on Biomedical Imaging (ISBI), 2016 IEEE 13th International Symposium on Biomedical Imaging (ISBI), 2016-04-13, Prague. 2016-04p. 1-4
IEEE
Resumen
3D reconstruction of the spine using biplanar X-rays remains approximate and requires human-machine interactions to adjust the position of important features such as vertebral corners and endplates. The purpose of this ...Leer más >
3D reconstruction of the spine using biplanar X-rays remains approximate and requires human-machine interactions to adjust the position of important features such as vertebral corners and endplates. The purpose of this study is to develop a method to extract automatically the accurate position of lumbar vertebrae posterior corners. In the proposed method we select corner point candidates from an initial edge map. A dedicated pipeline is designed to discard unwanted candidates, involving polyline simplification, curvature thresholding and multiscale Haar filtering. Ultimately, we use a priori knowledge derived from an initial 3D spine model to define search areas and select the final corner points. The framework was tested on 21 biplanar X-rays from scoliotic children. Corner positions are compared with manual selections by two experts. The results report a localization accuracy between 0.7 and 1.6 mm, comparable to manual expert variability.< Leer menos
Palabras clave
Biomedical Imaging
Image Analysis
Spine
Orígen
Importado de HalCentros de investigación