Low background germanium planar detector for gamma-ray spectrometry
REYSS, J.-L.
Laboratoire des Sciences du Climat et de l'Environnement [Gif-sur-Yvette] [LSCE]
Océan et Interfaces [OCEANIS]
< Leer menos
Laboratoire des Sciences du Climat et de l'Environnement [Gif-sur-Yvette] [LSCE]
Océan et Interfaces [OCEANIS]
Idioma
en
Article de revue
Este ítem está publicado en
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2011, vol. 634, p. 64-70
Elsevier
Resumen en inglés
A new ultra-low background planar germanium spectrometer has been developed. The planar geometry improves the sensitivity and energy resolution below 600 keV. The integral background counting rate in the Laboratoire ...Leer más >
A new ultra-low background planar germanium spectrometer has been developed. The planar geometry improves the sensitivity and energy resolution below 600 keV. The integral background counting rate in the Laboratoire Souterrain de Modane (4800 m water equivalent) in the energy range from 20 to 1500 keV for the planar Ge (mass=800 g) is 140 count/day. After 40 days of statistics, the background counting rates for all expected single lines are below 0.5 count/day with the exception of 210Pb(46-keV line) which was measured to be (1.76±0.25) count/day. Monte Carlo simulations have been performed to explain the origin of the remaining background and to calculate the detection efficiencies. Sensitivities around 1 mBq/kg are obtained within few days of statistics for 226Ra and 228Th. The main achievement is the high sensitivities for 210Pb (46-keV line) and 238U (234Th: 63 and 93 keV lines). For an aluminium sample (mass=1 kg) the limits obtained in 15 days are 210Pb<9 mBq/kg and 238U<3mBq/kg< Leer menos
Palabras clave en inglés
gamma-ray spectrometry
Low-background
Monte Carlo simulation