High-precision efficiency calibration of a high-purity co-axial germanium detector
Idioma
en
Article de revue
Este ítem está publicado en
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2015, vol. 776, p. 34-44
Elsevier
Resumen en inglés
A high-purity co-axial germanium detector has been calibrated in efficiency to a precision of about 0.15% over a wide energy range. High-precision scans of the detector crystal and gamma-ray source measurements have been ...Leer más >
A high-purity co-axial germanium detector has been calibrated in efficiency to a precision of about 0.15% over a wide energy range. High-precision scans of the detector crystal and gamma-ray source measurements have been compared to Monte-Carlo simulations to adjust the dimensions of a detector model. For this purpose, standard calibration sources and short-lived on-line sources have been used. The resulting efficiency calibration reaches the precision needed e.g. for branching ratio measurements of super-allowed beta decays for tests of the weak-interaction standard model.< Leer menos
Palabras clave en inglés
Super-allowed beta transitions
Monte-Carlosimulations
Gamma-ray spectroscopy
Germanium detector